Sensors

Sensor/Instrument catalog theory circuit setup parts calibration
wave gage     Click To Preview setup parts  
pressure sensors Omega        
Temperature probes Omega   Click To Preview   OL-703  
pH     pH Circuit setup   calibration
Dissolved Oxygen YSI theory Click To Preview setup   calibration
Load cells Omega          
Turbidity Sensors HFScientific     setup    

Sensor Voltage Ranges

  Part # power supply (V) Voltage range  Corresponding units Alternate units
Sensor     Vmin Vmax Vmin Vmax Vmax
7 kPa pressure PX26-001DV 10 0 16.7 mV  0 kPa 7 kPa 70 cm water
200 kPa pressure PX26-0030DV 10 0 100 mV 0 kPa 200 kPa 20 m water
100 kPa pressure PX136-015GV  10 0 100 mV 0 kPa 100 kPa 10 m water
Temperature OL-703 +5, -5 -200 200 mV 0 °C 40 °C  
Dissolved Oxygen 5331 +15, -15 0 200 mV 0 mg/L DO 10 mg/L DO  
pH

various

+15, -15 -420 420 mV 14 pH 0 pH  
  -200 200 mV 3.7 pH 10.3 pH  
Wave gages     -10  10 V based on potentiometer settings
Turbidity   5 0 5 V 0 NTU 1000 NTU  
Load Cells LCL-113G 10 0 20 mV   113 g  

For Instruments with LabVIEW software control see Software.

Ethernet Cable for wiring sensors

There are two standards, so check the RJ-45 plug to see which standard the cable uses before wiring the sensor! Failure to check this could cause damage to the sensor

T-568A Straight-Through Ethernet Cable

color code voltage
white/green S-
green S+
white/orange ground
blue -5 V
white/blue +5 V
orange +10V
white/brown -15 V
brown +15 V

The T-568A standard is supposed to be used in new network installations. Most off-the-shelf Ethernet cables are still of the T-568B standard; however, it makes absolutely no functional difference in which you choose.

T-568B Straight-Through Ethernet Cable

color code voltage
white/orange S-
orange S+
white/green ground
blue -5 V
white/blue +5 V
green +10V
white/brown -15 V
brown +15 V